monopolar mode- monopolar cutting
Monopolar cutting mode includes pure cut, blend 1, blend 2, blend 3. The output power can be adjustable.
Pure cut: cut the tissue cleanly and accurately without coagulation.
blend 1: Use when the cutting speed is slightly slow and a small amount of hemostasis is required.
blend 2: Compared with blend 1, it is used when the cutting speed is slightly slower and the better hemostatic effect is needed.
Monopolar modes- monopolar coagulation
Monopolar coagulation mode includes spray coagulation, forced coagulation, and the output power is adjustable.
forced coagulation: It’s non-contact coagulation. The output threshold voltage is lower than spray coagulation.
It’s suitable for coagulation in a small area.
pray coagulation: high-efficiency coagulation without contact surface. The coagulation depth is shallow.
The tissue is removed by evaporation. It usually use a Blade or ball electrode for coagulation.
Bipolar mode
The bipolar mode includes onlystandard mode. The output power is adjustable.
Standard mode: It’s suitable for most bipolar applications. Keep the low voltage to prevent sparks.
A new generation of intelligent tissue density immediate feedback technology
The device uses a new generation of intelligent tissue density immediate feedback technology. According to the different impedance of the tissue, the output is intelligently controlled to achieve the effect of minimally invasive tissue cutting, ensuring the output of ideal power, and reducing the thermal damage of the cut tissue.
Functional self-test
After each turn-on, the high-frequency electrosurgical unit will immediately execute a self-test procedure. Once the internal abnormality of the system is found and the self-test fails, the current output will be automatically cut off immediately. This ensures that the ES-100V generator is always in good working condition and performance. During the self-test, it is also tested whether the connected accessories are functioning normally.